Use when after peak detection when you have a table of detected peaks
Scanned 9/12/2026
Install to Claude Code
npx -y skills add HolobiomicsLab/asb-skill-collections --skill peak-m-z-recalibration --agent claude-codeInstalls into .claude/skills of the current project.
Are you the author of Peak M Z Recalibration?
Add the live security badge to your README — it updates automatically with every re-scan.
[](https://www.skillsdirectory.com/skills/holobiomicslab-peak-m-z-recalibration)More formats (shields.io, HTML) on the badges page.
---
name: peak-m-z-recalibration
description: Use when after peak detection when you have a table of detected peaks
with m/z values and need to improve mass accuracy for downstream annotation.
license: CC-BY-4.0
metadata:
edam_operation: http://edamontology.org/operation_3632
edam_topics:
- http://edamontology.org/topic_0121
- http://edamontology.org/topic_3370
tools:
- IDSL.IPA
- R
techniques:
- LC-MS
license_tier: restricted
provenance_tier: literature
derived_from:
- doi: 10.1021/acs.jproteome.2c00120
title: IDSL.IPA
evidence_spans:
- '**Intrinsic Peak Analysis (IPA)** by the [**Integrated Data Science Laboratory
for Metabolomics and Exposomics (IDSL.ME)**](https://www.idsl.me) is a light-weight
R package'
- light-weight R package
claims: []
provenance:
collection: https://w3id.org/holobiomicslab/asb-skill/collection/metabolomics/v2
assembled_by: scripts/collect_metabolomics_collection.py
sources:
- build: coll_idsl_ipa_cq
doi: 10.1021/acs.jproteome.2c00120
title: IDSL.IPA
dedup_kept_from: coll_idsl_ipa_cq
schema_version: 0.2.0
attribution:
generator: AgenticScienceBuilder
original_doi: 10.1021/acs.jproteome.2c00120
all_source_dois:
- 10.1021/acs.jproteome.2c00120
zenodo_doi: 10.5281/zenodo.20794027
curators: []
promoter: Louis-Félix Nothias
sponsor: CNRS & Université Côte d'Azur
---
# peak-m/z-recalibration
> **License: restricted** — no clear open-source license detected for the underlying tool; verify licensing before commercial use or redistribution. <!-- asb-license-banner -->
## Summary
Recursive mass correction refines m/z accuracy of detected LC/HRMS peaks by recalibrating their mass-to-charge coordinates using internal reference standards or lock masses. This step is applied post-detection to improve mass accuracy metrics across large untargeted metabolomics datasets.
## When to use
Apply this skill after peak detection when you have a table of detected peaks with m/z values and need to improve mass accuracy for downstream annotation. Use it in untargeted LC/HRMS population-scale studies (n > 500) where systematic mass drift or calibration error may affect m/z precision across multiple batches or acquisitions.
## When NOT to use
- Input peaks are already from a targeted method with locked or fixed m/z windows — recursive correction may introduce unnecessary noise.
- Untargeted data were acquired on low-resolution instruments (nominal mass accuracy) — correction tolerances and reference lock mass precision may not justify the algorithm.
- No internal reference standards or endogenous markers are available in the dataset — the algorithm requires calibration anchors to function.
## Inputs
- Detected peaks table (from peak detection stage) containing m/z values, retention times, and peak properties
- Mass spectrometry raw data (mzXML, mzML, or netCDF format) or extracted ion chromatogram (EIC) data
- Internal reference standards or lock mass m/z values (optional; endogenous markers may substitute)
## Outputs
- Corrected peak table with recalibrated m/z values
- Mass accuracy metrics (pre- and post-correction ppm error, calibration statistics)
- Calibration residual data for quality assessment
## How to apply
Load the detected peaks table containing m/z values and peak properties from the prior peak detection stage. Apply the recursive mass correction algorithm implemented in IDSL.IPA, which recalibrates m/z coordinates using internal reference standards (lock masses) or endogenous reference markers available in the dataset. The algorithm iteratively refines mass calibration to minimize systematic error. Generate a corrected peak table with updated m/z coordinates and compare pre- and post-correction m/z accuracy metrics (e.g., ppm error distribution, mass calibration residuals) to verify improvement. Export the final corrected peak dataset for downstream retention time correction and peak annotation workflows.
## Related tools
- **IDSL.IPA** (Provides the recursive mass correction algorithm and integration with full LC/HRMS peak processing pipeline (EIC generation, peak detection, property evaluation, retention time correction, and annotation)) — https://github.com/idslme/IDSL.IPA
- **R** (Language and runtime environment for executing IDSL.IPA package and recursive mass correction workflow)
## Examples
```
library(IDSL.IPA); IPA_workflow("path/to/IPA_parameters.xlsx")
```
## Evaluation signals
- Pre- and post-correction m/z accuracy distribution: post-correction ppm error should be narrower and centered closer to zero than pre-correction values.
- Calibration residuals: systematic mass drift should be minimized; residuals should not show batch or time-dependent patterns.
- Mass calibration statistics (e.g., RMSE, max error): quantitative improvement in mass precision across the m/z range.
- Downstream peak alignment and annotation quality: correctly recalibrated m/z values should improve alignment rates and reduce false monoisotopic peak assignments.
- Isotope pattern coherence: m/z differences between detected isotopologue pairs (e.g., ¹²C/¹³C) should match theoretical mass defect after correction.
## Limitations
- Recursive mass correction depends on availability and accuracy of internal reference standards or lock masses; if reference m/z values are themselves miscalibrated, the algorithm may propagate error.
- Performance in multi-batch studies requires endogenous reference markers (retention time correction anchors) to be present and reliably detected across all samples; sparse or variable reference detection can degrade correction fidelity.
- The algorithm is designed for high-resolution mass spectrometry (HRMS) data; application to nominal-mass or low-resolution data may not yield meaningful improvements.
- No changelog was found in the repository documentation, limiting visibility into algorithm refinements or known issues across IDSL.IPA versions.
## Evidence
- [other] IDSL.IPA includes a recursive mass correction algorithm as part of its suite of peak processing algorithms, applied after peak detection to refine mass accuracy of detected peaks.: "recursive mass correction algorithm as part of its suite of peak processing algorithms, applied after peak detection to refine mass accuracy"
- [other] Load detected peaks table containing m/z values and peak properties from prior detection step. Apply recursive mass correction algorithm as implemented in IDSL.IPA to recalibrate m/z values using internal reference standards or lock masses.: "Load detected peaks table containing m/z values and peak properties from prior detection step. Apply recursive mass correction algorithm as implemented in IDSL.IPA to recalibrate m/z values using"
- [readme] IDSL.IPA is a suite of new algorithms covering extracted ion chromatogram (EIC) candidate generation, peak detection, peak property evaluation, recursive mass correction, retention time correction across multiple batches and peak annotation.: "IDSL.IPA is a suite of new algorithms covering extracted ion chromatogram (EIC) candidate generation, peak detection, peak property evaluation, recursive mass correction, retention time correction"
- [readme] extracts peaks for organic small molecules from untargeted liquid chromatography high resolution mass spectrometry (LC/HRMS) data in population scale projects: "extracts peaks for organic small molecules from untargeted liquid chromatography high resolution mass spectrometry (LC/HRMS) data in population scale projects"
- [readme] Retention time correction using endogenous reference markers for multi-batch large scale studies: "Retention time correction using endogenous reference markers for multi-batch large scale studies"