Use when you have loaded a Bruker Solarix transient file (.d format with
Scanned 9/12/2026
Install to Claude Code
npx -y skills add HolobiomicsLab/asb-skill-collections --skill mass-spectrum-averaging-and-processing --agent claude-codeInstalls into .claude/skills of the current project.
Are you the author of Mass Spectrum Averaging And Processing?
Add the live security badge to your README — it updates automatically with every re-scan.
[](https://www.skillsdirectory.com/skills/holobiomicslab-mass-spectrum-averaging-and-processing)More formats (shields.io, HTML) on the badges page.
---
name: mass-spectrum-averaging-and-processing
description: Use when you have loaded a Bruker Solarix transient file (.d format with
.ser or .fid content) and need to generate a processed mass spectrum for peak picking
and molecular formula annotation.
license: CC-BY-4.0
metadata:
edam_operation: http://edamontology.org/operation_3627
edam_topics:
- http://edamontology.org/topic_0121
- http://edamontology.org/topic_3520
tools:
- CoreMS
- numpy
- matplotlib
techniques:
- LC-MS
license_tier: open
provenance_tier: literature
derived_from:
- doi: 10.5281/zenodo.14009575
title: corems
evidence_spans:
- from corems.transient.input.brukerSolarix import ReadBrukerSolarix
- '**CoreMS** is a comprehensive mass spectrometry framework for software development
and data analysis of small molecules analysis.'
- import numpy as np
- from matplotlib import pyplot
claims: []
provenance:
collection: https://w3id.org/holobiomicslab/asb-skill/collection/metabolomics/v2
assembled_by: scripts/collect_metabolomics_collection.py
sources:
- build: coll_corems_cq
doi: 10.5281/zenodo.14009575
title: corems
dedup_kept_from: coll_corems_cq
schema_version: 0.2.0
attribution:
generator: AgenticScienceBuilder
original_doi: 10.5281/zenodo.14009575
all_source_dois:
- 10.5281/zenodo.14009575
zenodo_doi: 10.5281/zenodo.20794027
curators: []
promoter: Louis-Félix Nothias
sponsor: CNRS & Université Côte d'Azur
---
# mass-spectrum-averaging-and-processing
## Summary
Apply transient averaging and windowing functions (e.g., Hanning apodization) to Bruker FT-ICR raw transient data, followed by zero-filling and Fourier transformation, to produce a high-quality mass spectrum with reduced noise and improved peak definition. This skill is essential for converting raw time-domain transient data into interpretable frequency-domain mass spectra.
## When to use
You have loaded a Bruker Solarix transient file (.d format with .ser or .fid content) and need to generate a processed mass spectrum for peak picking and molecular formula annotation. Use this skill when raw transient data requires noise reduction, improved peak resolution, and frequency-domain conversion to enable downstream analysis such as Kendrick classification or molecular formulae assignment.
## When NOT to use
- Input is already a centroided mass spectrum or peak list; re-processing would introduce artifacts.
- Data originates from a vendor already providing processed spectra (e.g., Bruker CompassXtract output); re-processing raw transients is redundant.
- Transient file is corrupt or truncated; processing will fail or produce spurious peaks.
## Inputs
- Bruker Solarix transient file (.d directory with .ser or .fid files)
- MSParameters configuration object specifying apodization, zero-fill, and averaging parameters
## Outputs
- Processed mass spectrum object with detected peaks (m/z, abundance, resolving power)
- Peak list with m/z values, abundances, and associated metadata
## How to apply
Import the transient using CoreMS's ReadBrukerSolarix class and configure MSParameters to specify apodization window function (e.g., Hanning), zero-fill factor (e.g., 1), and averaging settings. Process the transient data through CoreMS's transient processing pipeline, which applies the selected apodization window, performs zero-filling, and computes the magnitude-mode Fourier transform to generate the final mass spectrum. Verify output by inspecting the resulting peak count, m/z range, and abundance distribution; peaks should span the expected m/z window (e.g., m/z 155–1000 for SRFA ESI-negative data) with noise floor clearly below the lowest detected peaks.
## Related tools
- **CoreMS** (Core framework providing ReadBrukerSolarix transient import, apodization window application, zero-filling, Fourier transform, and peak detection algorithms) — https://github.com/EMSL-Computing/CoreMS
- **numpy** (Underlying numerical library for FFT and array operations during transient processing)
- **matplotlib** (Visualization of processed mass spectrum and peak picking results)
## Examples
```
from corems.transient.input.brukerSolarix import ReadBrukerSolarix; lcms = ReadBrukerSolarix('tests/tests_data/ftms/ESI_NEG_SRFA.d'); lcms.transient_process()
```
## Evaluation signals
- Peak count matches literature or expected range for the sample (e.g., ~10,588 peaks for SRFA ESI-NEG); order-of-magnitude deviation suggests processing error or wrong parameters.
- m/z range of detected peaks spans expected molecular weight window (e.g., m/z 155–1000 for natural organic matter); gaps or truncation indicate calibration or processing failure.
- Abundance distribution shows clear separation between noise floor and detected peaks; if lowest-abundance peaks approach noise floor or peaks appear uniformly distributed, windowing or threshold parameters need adjustment.
- Resolving power values are consistent with instrument specification and transient time; resolving power should scale with transient length and magnetic field strength.
- No spurious high-abundance peaks at m/z extremes or with unrealistic isotope patterns; such artifacts indicate phase correction, zero-fill, or apodization errors.
## Limitations
- Apodization window choice (Hanning vs. others) affects peak shape and sensitivity; Hanning reduces spectral noise but may broaden peaks and reduce mass resolving power compared to other windows.
- Zero-filling factor impacts spectral appearance and computational cost; insufficient zero-fill (e.g., factor=1) may not fully resolve closely spaced peaks, while excessive zero-fill adds computational overhead without improving mass accuracy.
- Processing assumes transient data is in magnitude-mode FT format; phase-corrected or other non-magnitude representations will produce incorrect spectra.
- Noise threshold settings (relative_abundance, log, signal_noise) must be tuned per sample; inappropriate thresholds will either lose low-abundance peaks or retain excessive noise.
- Phase correction and frequency axis calibration are prerequisites; uncorrected phase errors or miscalibrated frequency offsets will distort peak positions and shapes.
## Evidence
- [other] Using Hanning apodization with zero fill settings on ESI_NEG_SRFA.d produces a mass spectrum with 10588 detected m/z peaks spanning from m/z 155.87 to m/z 999.66.: "Using Hanning apodization with zero fill settings on ESI_NEG_SRFA.d produces a mass spectrum with 10588 detected m/z peaks spanning from m/z 155.87 to m/z 999.66."
- [other] Import the ReadBrukerSolarix class from CoreMS and load the ESI_NEG_SRFA.d transient file. Configure transient processing parameters to apply Hanning apodization window function. Process the transient data to generate the mass spectrum with averaging applied.: "Import the ReadBrukerSolarix class from CoreMS and load the ESI_NEG_SRFA.d transient file. Configure transient processing parameters to apply Hanning apodization window function. Process the"
- [readme] Apodization, Zerofilling, and Magnitude mode FT: "Apodization, Zerofilling, and Magnitude mode FT"
- [readme] from corems.transient.input.brukerSolarix import ReadBrukerSolarix: "from corems.transient.input.brukerSolarix import ReadBrukerSolarix"