Compute the surface loading a vacuum-probe or solvent-rinse cleanliness sample actually reports under ECSS-Q-ST-70-50C. Use when hardware surfaces have been sampled for monitoring, each sample carries a witness blank and a sampled area, and the record is either an evaporated rinsate aliquot or a probe stroke pattern. Scales the aliquot back to the whole collected volume, subtracts the blank, divides out the collection recovery of the method, turns a stroke pattern into the area it truly swept...
Scanned 9/27/2026
Install to Claude Code
npx -y skills add ashfordeOU/aero-agent-skills --skill q7050-vacuum-and-rinse-methods --agent claude-codeInstalls into .claude/skills of the current project.
Are you the author of Q7050 Vacuum And Rinse Methods?
Add the live security badge to your README — it updates automatically with every re-scan.
[](https://www.skillsdirectory.com/skills/ashfordeou-q7050-vacuum-and-rinse-methods)More formats (shields.io, HTML) on the badges page.
---
name: q7050-vacuum-and-rinse-methods
description: "Compute the surface loading a vacuum-probe or solvent-rinse cleanliness sample actually reports under ECSS-Q-ST-70-50C. Use when hardware surfaces have been sampled for monitoring, each sample carries a witness blank and a sampled area, and the record is either an evaporated rinsate aliquot or a probe stroke pattern. Scales the aliquot back to the whole collected volume, subtracts the blank, divides out the collection recovery of the method, turns a stroke pattern into the area it truly swept rather than the area it covered, holds a net the balance cannot resolve to a bounded statement, and grades the method detection limit against the reporting limit. Trigger: ecss, q-st-70-50, surface-particulate-sampling, solvent-rinse-nvr, vacuum-probe-stroke-coverage, witness-blank-subtraction, sampling-recovery-fraction, nvr-area-density."
license: Apache-2.0
compliance: STANDARDS-REF
standards:
- id: ecss
reference-only: true
gated: false
domain: space-systems
pack: space-systems
compatibility: "agentskills.io SKILL.md; any SKILL.md host (Claude Code, Hermes, OpenClaw)"
metadata:
domain: space-systems
subdomain: ecss
tags: [ecss, q-st-70-cleanliness-monitoring-scope, q7050-vacuum-and-rinse-methods, surface-particulate-sampling, solvent-rinse-nvr, vacuum-probe-stroke-coverage, witness-blank-subtraction, sampling-recovery-fraction]
version: 0.1.0
author: Aero Agent Skills
---
# ECSS Cleanliness Monitoring — Vacuum and Rinse Sampling (space-systems/ecss/q7050-vacuum-and-rinse-methods)
Use when the task is the surface-sampling step of ECSS-Q-ST-70-50C
cleanliness monitoring — taking a vacuum-probe or solvent-rinse record
off a piece of hardware and turning it into a loading per unit area that
a contamination budget can be built on.
## Domain quick reference
- A sampling result is a mass or a count divided by an area, and the
area is the half of that ratio most often wrong. A rinse samples the
wetted area, which is not the projected area of the part; a probe
samples the area its strokes swept, which is less than the area the
pattern covered because adjacent strokes are deliberately overlapped
so no lane is missed.
- A rinse rarely evaporates the whole rinsate. One aliquot is taken to
a tared dish, so the weighed residue belongs to that aliquot and has
to be scaled by the collected volume over the aliquot volume before it
belongs to the surface.
- The witness blank is the solvent, the glassware and the handling, run
through the same path with no hardware in it. Subtracting it is not
optional: on a clean surface the blank can be the larger of the two
numbers, and a sample under its own blank past the weighing noise says
the rinsate was contaminated, not that the surface was.
- Neither method lifts everything. The recovery fraction of the method
on that surface finish divides into the collected mass to give the
surface loading; leaving it out reports the sampler's efficiency as if
it were the hardware's cleanliness.
- The detection limit is a property of the sample, not of the balance
alone: the same readability over a smaller area, or behind a larger
dilution, cannot reach the same reporting limit. A sample whose floor
sits above the programme's reporting limit answers nothing, however
clean the result looks.
## Workflow
1. Decide the method the surface can carry — accessibility, solvent
compatibility, whether rinsate can be collected, whether the surface
tolerates a probe — and refuse a surface that supports neither rather
than forcing a sample it cannot give.
2. Establish the area: the wetted area for a rinse; for a probe, the
swept area computed from stroke count, probe width, stroke length and
the overlap fraction between strokes.
3. Subtract the witness blank from the weighed residue and categorize
the net as quantified, sub-floor, or blank-dominated against the
balance floor.
4. Scale a rinse residue from the aliquot to the whole collected volume.
5. Divide by the collection recovery of the method so the figure is
surface loading, not collected loading.
6. Express the result per unit area and compute the detection limit the
same area, recovery and dilution imply.
7. Report each sample with its status, its loading and its floor, and
raise a finding for a blank-dominated sample, a loading above the
reporting limit, or a floor that cannot reach that limit.
## Pitfalls
- Reporting the aliquot residue as the sample mass. A 100 mL aliquot of
a 500 mL rinsate understates the surface by a factor of five, and the
result still looks like a plausible number.
- Counting the probe pattern area instead of the swept area. Ten
strokes at half overlap sweep barely more than half the lanes their
pattern spans, so the density comes out low by nearly a factor of two.
- Dropping the witness blank because it was small. It is small compared
with a dirty surface and comparable with a clean one, which is exactly
the region where the monitoring decision is being made.
- Treating a sub-floor net as a measurement. The balance cannot separate
it from zero, so it is a bound; carried into a budget as a small
positive number it gives the budget false precision.
- Accepting a sample whose detection limit sits above the reporting
limit. A non-detect there is not a clean surface, it is an unanswered
question, and enlarging the area or reducing the dilution is the fix.
## Behavior contract (gate 3)
The method selection, swept-area geometry, aliquot scaling, blank
subtraction, recovery correction, area-density conversion and
detection-limit grading are exercised by the gate 3 contract test:
scripts/test_q7050_vacuum_and_rinse_methods.py against
scripts/q7050_vacuum_and_rinse_methods_logic.py (stdlib unittest,
offline). Run:
python3 scripts/test_q7050_vacuum_and_rinse_methods.py
## Compliance
- ECSS standards are freely downloadable (ESA); cite the source and
paraphrase per standards-map.yaml.
- compliance: STANDARDS-REF, gated: false.