Use when verify the ambient radiated-emission background of an EMC facility under ECSS-E-ST-20-07C clause 5.2.2.3: confirm the baseline run had the unit-under-test unpowered while the support-equipment stayed operating behind a closed enclosure door, sweep the recorded ambient-noise-floor against the applicable radiated-emission-limit, categorize every frequency as compliant, marginal or exceeding once the required headroom is applied, flag narrowband ambient signals that must be documented, ...
Scanned 9/27/2026
Install to Claude Code
npx -y skills add ashfordeOU/aero-agent-skills --skill e2007-ambient-radiated-level-check --agent claude-codeInstalls into .claude/skills of the current project.
Are you the author of E2007 Ambient Radiated Level Check?
Add the live security badge to your README — it updates automatically with every re-scan.
[](https://www.skillsdirectory.com/skills/ashfordeou-e2007-ambient-radiated-level-check)More formats (shields.io, HTML) on the badges page.
---
name: e2007-ambient-radiated-level-check
description: "Use when verify the ambient radiated-emission background of an EMC facility under ECSS-E-ST-20-07C clause 5.2.2.3: confirm the baseline run had the unit-under-test unpowered while the support-equipment stayed operating behind a closed enclosure door, sweep the recorded ambient-noise-floor against the applicable radiated-emission-limit, categorize every frequency as compliant, marginal or exceeding once the required headroom is applied, flag narrowband ambient signals that must be documented, check the swept band for coverage gaps, and decide whether a later measured level is attributable to the unit or is ambient-limited. Trigger: ecss, e-st-20-07c, ambient-radiated-level, background-noise-floor, eut-off-baseline, support-equipment-operating, radiated-emission-limit-headroom, narrowband-ambient-signal, emc-facility-baseline."
license: Apache-2.0
compliance: STANDARDS-REF
standards:
- id: ecss
reference-only: true
gated: false
domain: space-systems
pack: space-systems
compatibility: "agentskills.io SKILL.md; any SKILL.md host (Claude Code, Hermes, OpenClaw)"
metadata:
domain: space-systems
subdomain: ecss
tags: [ecss, e-st-20-electrical-scope, e2007-ambient-radiated-level-check, ambient-radiated-level, background-noise-floor, eut-off-baseline, radiated-emission-limit-headroom, narrowband-ambient-signal, emc-facility-baseline]
version: 0.1.0
author: Aero Agent Skills
---
# ECSS EMC Facility — Ambient Radiated-Level Check (space-systems/ecss/e2007-ambient-radiated-level-check)
Use when the task is the ambient radiated background baseline of
ECSS-E-ST-20-07C clause 5.2.2.3 -- recording and grading the
radiated-emission floor of the facility with the unit-under-test
unpowered and the support-equipment running, so that every later
emission reading can be attributed to the unit rather than to the
room.
## Domain quick reference
- The ambient run is not "the empty room". The unit-under-test is
unpowered, but the support-equipment, harness, stimulation gear and
facility services stay in exactly the configuration the graded run
will use, with the enclosure door closed. An ambient run taken with
the support-equipment switched off understates the floor and
invalidates every later attribution.
- The ambient sweep must reuse the graded run's measurement settings:
same antenna polarization, same standoff, same receiver bandwidth.
A narrower receiver bandwidth lowers the apparent floor without
lowering the real one, so a mismatched bandwidth is a defect in the
baseline, not a favourable result.
- Grading is by headroom, the difference between the applicable
radiated-emission-limit and the recorded ambient level at that
frequency. A required headroom (conventionally 6 dB) separates a
usable baseline from one where the room itself is close enough to
the limit that a compliant unit could still read as failing.
Headroom at or above the requirement is compliant; positive but
short of it is marginal and must be recorded as a limitation;
headroom at or below zero is an ambient exceedance and the
frequency cannot be graded at that facility until it is resolved.
- A narrowband ambient signal is a discrete peak that stands well
above its local neighbourhood -- a broadcast carrier, a local
oscillator, a nearby transmitter. Clause 5.2.2.3 practice allows a
known narrowband ambient to be documented and carried, provided it
is identified and shown not to originate from the unit; a broadband
floor that is simply high cannot be carried this way.
- Coverage matters as much as level. The recorded points must span
the whole declared band, and consecutive points must not step
further apart than the declared maximum ratio, otherwise a narrow
ambient peak can hide between two samples.
- Attribution rules a later graded reading: a measured level well
above the ambient is dominated by the unit, a measured level equal
to the ambient is ambient-limited and carries no information about
the unit, and the band between them is indeterminate -- the reading
proves neither compliance nor a finding.
## Workflow
1. Validate the baseline configuration: unit-under-test unpowered,
support-equipment powered, enclosure door closed, receiver
bandwidth positive and matching the graded run, recognized antenna
polarization, positive standoff. Reject the run rather than grading
a configuration that cannot support attribution.
2. Validate the sweep: at least one point, strictly increasing
frequencies, positive frequencies, an ambient level and an
applicable limit at every point.
3. For each point compute headroom = limit - ambient, and categorize
it as compliant, marginal or exceedance against the required
headroom. Absorb representation error at the boundary with a
named decibel tolerance; never lower the required headroom to make
a boundary point pass.
4. Identify narrowband ambient peaks by comparing each interior point
against its two neighbours; report them for documentation.
5. Check band coverage: first point at or below the band start, last
point at or above the band stop, and no step ratio above the
declared maximum.
6. Aggregate: the worst-case frequency, the counts per category, the
coverage gaps, and a verdict that is usable only when no exceedance
and no coverage gap remains.
7. When grading a later reading, apply the attribution rule to the
measured/ambient pair before drawing any conclusion from it.
## Pitfalls
- Running the ambient sweep with the support-equipment off because
"only the unit matters" -- the resulting floor is not the floor the
graded run will see.
- Reading "ambient below the limit" as a usable baseline. Below the
limit but inside the required headroom is a marginal frequency and
has to be carried as a limitation on the measurement.
- Grading a reading that sits on top of the ambient. An emission equal
to the ambient is ambient-limited; reporting it as a unit emission
invents a finding, and reporting it as a pass invents compliance.
- Sampling coarsely and declaring the band clean. A narrowband carrier
between two widely spaced points is invisible to the sweep and will
reappear in the graded run.
- Narrowing the receiver bandwidth for the ambient run only. The floor
drops, the graded run does not, and every later comparison is wrong.
## Behavior contract (gate 3)
The configuration-validation, headroom-categorization, narrowband
identification, coverage and attribution logic is exercised by the
gate 3 contract test:
scripts/test_e2007_ambient_radiated_level_check.py against
scripts/e2007_ambient_radiated_level_check_logic.py (stdlib unittest,
offline). Run: python3 scripts/test_e2007_ambient_radiated_level_check.py
## Compliance
- ECSS standards are freely downloadable (ESA); cite the source and
paraphrase per standards-map.yaml.
- compliance: STANDARDS-REF, gated: false.