Use when verify that a secondary-emission measurement-facility holds valid calibration under ECSS-E-ST-20-01C clause 9.5.2: check every instrumented channel -- beam-current, collector-current, beam-energy, base-pressure, sample-temperature -- was calibrated before the run and still inside its validity interval, confirm each calibration is traceable to a recognized reference-standard, compare pre-run against post-run check readings for instrument-drift beyond the stated tolerance, combine the ...
Scanned 9/27/2026
Install to Claude Code
npx -y skills add ashfordeOU/aero-agent-skills --skill e2001-measurement-facility-calibration --agent claude-codeInstalls into .claude/skills of the current project.
Are you the author of E2001 Measurement Facility Calibration?
Add the live security badge to your README — it updates automatically with every re-scan.
[](https://www.skillsdirectory.com/skills/ashfordeou-e2001-measurement-facility-calibration)More formats (shields.io, HTML) on the badges page.
---
name: e2001-measurement-facility-calibration
description: "Use when verify that a secondary-emission measurement-facility holds valid calibration under ECSS-E-ST-20-01C clause 9.5.2: check every instrumented channel -- beam-current, collector-current, beam-energy, base-pressure, sample-temperature -- was calibrated before the run and still inside its validity interval, confirm each calibration is traceable to a recognized reference-standard, compare pre-run against post-run check readings for instrument-drift beyond the stated tolerance, combine the per-channel uncertainty contributions by root-sum-square into a combined-standard-uncertainty and expand it with a declared coverage-factor, and confirm the calibration-record actually reached the customer with recipient, issue date and covered channels. Trigger: ecss, e-st-20-01c, measurement-facility-calibration, calibration-validity-interval, metrological-traceability, instrument-drift-check, calibration-record-delivery."
license: Apache-2.0
compliance: STANDARDS-REF
standards:
- id: ecss
reference-only: true
gated: false
domain: space-systems
pack: space-systems
compatibility: "agentskills.io SKILL.md; any SKILL.md host (Claude Code, Hermes, OpenClaw)"
metadata:
domain: space-systems
subdomain: ecss
tags: [ecss, e-st-20-electrical-scope, e2001-measurement-facility-calibration, measurement-facility-calibration, calibration-validity-interval, metrological-traceability, instrument-drift-check, calibration-record-delivery]
version: 0.1.0
author: Aero Agent Skills
---
# ECSS Multipaction — Measurement-Facility Calibration (space-systems/ecss/e2001-measurement-facility-calibration)
Use when the task is the clause 9.5.2 obligation of ECSS-E-ST-20-01C:
establishing that the facility used for a secondary-electron-emission
yield measurement was calibrated, that the calibration was still in
force when the run happened, and that the calibration results were put
in the customer's hands rather than filed in the laboratory.
## Domain quick reference
- Clause 9.5.2 carries two coupled duties. The first is metrological:
the measurement-facility is calibrated, and the calibration is real
-- dated, traceable, and still inside its validity interval on the
day of the run. The second is contractual: the calibration results
are made available to the customer. A facility can satisfy the first
and fail the second, and the clause is then unmet.
- Five channels carry the emission-yield result and are therefore
required: beam-current, collector-current, beam-energy,
base-pressure and sample-temperature. Further channels (chamber
residual-gas composition, stage position) are supplementary --
recorded, never counted as required. A channel name that matches
neither list is uncategorized and never discharges a required
channel.
- Validity is an interval, not a date. A channel calibrated on day D
with an interval of N days is in force for measurements on days D
through D+N inclusive; a run on D+N+1 is out of calibration, and a
run before D means the certificate postdates the data it is meant to
underwrite, which is an error rather than a finding.
- Traceability is the property that links a facility reading to a
recognized reference-standard through an unbroken chain. This leaf
accepts a declared chain to a national-metrology-institute, an
accredited-calibration-laboratory, or a certified transfer-standard.
A channel calibrated against an in-house instrument with no declared
chain is untraceable, whatever its paperwork says.
- Instrument-drift is checked by reading a stable reference before and
after the run. The relative drift is the change divided by the
pre-run reading, in percent, against a per-channel tolerance. A
pre-run reading of zero makes the relative drift undefined and is
rejected rather than divided through.
- The uncertainty budget combines independent per-channel standard
contributions by root-sum-square into a combined-standard-
uncertainty, then multiplies by a declared coverage-factor to report
an expanded uncertainty. The coverage-factor must be stated and
positive; reporting an expanded value without naming the factor
makes the number uninterpretable.
## Workflow
1. Categorize every declared channel as required, supplementary or
uncategorized, and reject a channel entry that is not a mapping or
carries no name.
2. For each channel, parse the calibration date and the measurement
date, reject a calibration dated after the run, and place the run
inside or outside the validity interval (inclusive at both ends).
3. Confirm each channel's declared traceability chain resolves to a
recognized reference-standard; flag any channel whose chain is
absent or unrecognized.
4. Where pre-run and post-run check readings exist, compute the
relative instrument-drift and compare it with the channel
tolerance, absorbing representation residue in the comparison so a
drift sitting exactly on the tolerance is not failed by rounding.
5. Combine the per-channel standard uncertainty contributions by
root-sum-square, compare the combined value with the facility
budget under the same tolerance discipline, and expand it with the
declared coverage-factor.
6. Check the delivery record: a named recipient, an issue date not
earlier than the latest calibration date, and coverage of every
required channel. Aggregate the metrological and delivery findings
-- the clause is met only when both lists are empty and no required
channel is missing.
## Pitfalls
- Reading "calibrated" as a boolean. A certificate with no date, no
interval or no traceability chain establishes nothing; the audit is
of the three attributes together, not of the word.
- Failing a run that falls exactly on the last day of the validity
interval. The interval is inclusive; an off-by-one here rejects
perfectly valid data and invites the laboratory to widen intervals
to compensate.
- Summing uncertainty contributions arithmetically. Independent
contributions combine in quadrature; a linear sum overstates the
combined-standard-uncertainty and can push a compliant facility over
its budget for no physical reason.
- Comparing a root-sum-square against its budget with a bare
inequality. A quadrature sum of exactly-budgeted contributions lands
a few units in the last place above the budget; absorb that in the
comparison, never by raising the budget.
- Treating an internal file copy as availability to the customer.
Clause 9.5.2 is discharged by delivery -- recipient, date, and the
channels covered -- not by the existence of a certificate somewhere
in the facility's archive.
- Letting a supplementary channel stand in for a required one. A
well-calibrated residual-gas channel does not compensate for an
uncalibrated collector-current channel.
## Behavior contract (gate 3)
The channel categorization, validity-interval placement, traceability
check, drift computation, uncertainty combination and delivery check
are exercised by the gate 3 contract test:
`scripts/test_e2001_measurement_facility_calibration.py` against
`scripts/e2001_measurement_facility_calibration_logic.py`
(stdlib unittest, offline). Run:
python3 scripts/test_e2001_measurement_facility_calibration.py
## Compliance
- ECSS standards are freely downloadable (ESA); cite the source and
paraphrase per standards-map.yaml.
- compliance: STANDARDS-REF, gated: false.