Use when when you need to assess whether retention times measured on a given LC-MS run follow the expected linear relationship defined by iRT peptide standards (e.g., Pierce or Biognosys iRT peptides).
Scanned 9/12/2026
Install to Claude Code
npx -y skills add HolobiomicsLab/asb-skill-collections --skill irt-peptide-calibration-and-scoring --agent claude-codeInstalls into .claude/skills of the current project.
Are you the author of Irt Peptide Calibration And Scoring?
Add the live security badge to your README — it updates automatically with every re-scan.
[](https://www.skillsdirectory.com/skills/holobiomicslab-irt-peptide-calibration-and-scoring-77c89980)More formats (shields.io, HTML) on the badges page.
---
name: irt-peptide-calibration-and-scoring
description: Use when when you need to assess whether retention times measured on a given LC-MS run follow the expected linear relationship defined by iRT peptide standards (e.g., Pierce or Biognosys iRT peptides).
license: CC-BY-4.0
metadata:
edam_operation: http://edamontology.org/operation_3435
edam_topics:
- http://edamontology.org/topic_0121
- http://edamontology.org/topic_3520
tools:
- rawrr
- RawFileReader
- R base (stats::lm)
techniques:
- LC-MS
derived_from:
- doi: 10.1101/2020.10.30.362533
title: rawrr
- doi: 10.1021/acs.jproteome.0c00866
title: ''
evidence_spans:
- rawrr::readSpectrum
- Our .NET 8.0 [@dotnet] precompiled wrapper methods are bundled, including the runtime, in the `r BiocStyle::Biocpkg('rawrr')` executable file
- The extracted information is written to a temporary location on the harddrive, read back into memory and parsed into `R` objects using RawFileReader API
- 'ThermoFisher.CommonCore dlls can be obtained through: https://github.com/thermofisherlsms/RawFileReader'
claims: []
provenance:
collection: https://w3id.org/holobiomicslab/asb-skill/collection/metabolomics/v2
assembled_by: scripts/collect_metabolomics_collection.py
sources:
- build: coll_rawrr
doi: 10.1101/2020.10.30.362533
title: rawrr
dedup_kept_from: coll_rawrr
schema_version: 0.2.0
attribution:
generator: AgenticScienceBuilder
original_doi: 10.1101/2020.10.30.362533
all_source_dois:
- 10.1101/2020.10.30.362533
- 10.1021/acs.jproteome.0c00866
zenodo_doi: 10.5281/zenodo.20794027
curators: []
promoter: Louis-Félix Nothias
sponsor: CNRS & Université Côte d'Azur
---
# iRT-peptide-calibration-and-scoring
## Summary
Validates retention time (RT) linearity and instrument calibration by fitting a linear regression model to extracted ion chromatograms (XICs) of known iRT peptide standards, using their annotated iRT scores as the independent variable. The R-squared value (target ≥0.99) quantifies the degree of linear RT behavior across the peptide set.
## When to use
When you need to assess whether retention times measured on a given LC-MS run follow the expected linear relationship defined by iRT peptide standards (e.g., Pierce or Biognosys iRT peptides). This is appropriate when you have extracted XICs for 11+ known iRT peptides from a raw mass spectrometry file and want to validate instrument RT calibration quality or establish a RT prediction model for peptide identification.
## When NOT to use
- When the input raw file does not contain iRT peptide standards or the peptides were not ionized/detected (XICs are empty or sparse).
- When fewer than 5–6 iRT peptides are confidently detected; the regression requires sufficient degrees of freedom and coverage of the RT range.
- When you are analyzing a non-Orbitrap or non-Thermo instrument; the rawrr package is specific to Thermo Fisher Scientific .NET assemblies and .raw files.
## Inputs
- Thermo Fisher Scientific .raw file (Orbitrap instrument data)
- List of known iRT peptide precursor m/z values
- Annotated iRT reference scores (e.g., Pierce iRT standard peptides)
- Mass tolerance parameter (e.g., 10 ppm)
## Outputs
- rawrrChromatogram object (fitted intensity traces)
- Vector of rtFittedAPEX values (observed RT at peak maximum for each peptide)
- Linear regression model object (lm class in R)
- R-squared value quantifying RT linearity
- Residual plot or diagnostic statistics
## How to apply
Extract precursor m/z values and measured retention times for each iRT peptide standard using the rawrr::readChromatogram() function with mass tolerance ~10 ppm and XIC filter type. Fit intensity traces stored in rawrrChromatogram objects to identify the retention time at maximum intensity (rtFittedAPEX) for each peptide. Construct a linear regression model lm(rtFittedAPEX ~ iRTscore) where iRTscore is the annotated reference score for each peptide. Report the resulting R-squared value; values ≥0.99 indicate highly linear RT behavior and good instrument calibration. Lower R-squared values may suggest drift, temperature variation, or column degradation.
## Related tools
- **rawrr** (R package providing direct access to Thermo Orbitrap .raw file data; used to read file headers, extract XICs, and retrieve spectrum metadata) — https://github.com/fgcz/rawrr
- **RawFileReader** (.NET assembly (C#) underlying rawrr; implements low-level chromatogram and spectrum reading from binary .raw files) — https://github.com/thermofisherlsms/RawFileReader
- **R base (stats::lm)** (Fits linear regression model to RT observations and iRT scores; computes R-squared and residuals)
## Examples
```
C <- rawrr::readChromatogram(rawfile, mass = iRTmz, tol = 10, type = "xic", filter = "ms"); m <- lm(rtFittedAPEX ~ iRTscore); summary(m)$r.squared
```
## Evaluation signals
- R-squared value is ≥0.99 for iRT regression, confirming highly linear RT behavior across the 11+ peptides.
- Residual plot shows no systematic trend (e.g., funnel or sine pattern), indicating homoscedasticity and model validity.
- All iRT peptides are detected (non-empty XICs) and rtFittedAPEX values span the full RT range of the run without outliers.
- Fitted model coefficients (intercept and slope) are stable across replicate measurements of the same raw file, indicating reproducibility.
- Root mean square error (RMSE) or mean absolute error (MAE) of predicted vs. observed RT is <0.5 min, depending on method-specific tolerance.
## Limitations
- The rawrr package and RawFileReader .NET assembly are specific to Thermo Fisher Scientific Orbitrap instruments and .raw files; not applicable to other vendor formats (e.g., Bruker .d, Waters .raw, AB Sciex .wiff).
- On Windows systems, the decimal symbol must be configured as '.' for proper data extraction; regional locale settings can cause parsing errors.
- The quality of the regression depends on confident detection and annotation of iRT peptides; missing or misidentified peptides (e.g., due to ionization suppression or coelution) will reduce R-squared.
- RT linearity assumes stable column conditions and temperature; significant drift or column degradation during the run can artificially lower R-squared even with good instrument calibration.
- The choice of mass tolerance (e.g., 10 ppm) and XIC extraction filter directly affects which scans are included; too loose a tolerance may capture co-eluting interferents, biasing rtFittedAPEX.
## Evidence
- [results] Extract precursor m/z and RT using rawrr::readChromatogram() with mass tolerance and XIC filter: "Extract precursor m/z values and retention times for known iRT peptide standards using rawrr::readChromatogram() with mass tolerance 10 ppm and XIC filter."
- [results] Fit intensity traces to extract rtFittedAPEX and construct linear regression model: "we extract the RTs at the maximum of the fitted intensity traces stored in the `rawrrChromatogram` object and fit a linear model"
- [results] R-squared of 0.9999 indicates highly linear RT behavior: "The corresponding R-squared indicates that the RTs behave highly linear"
- [intro] rawrr wraps RawFileReader .NET assembly for direct access to Orbitrap data: "rawrr wraps the functionality of the RawFileReader .NET assembly"
- [intro] Gap in R ecosystem for raw data reading; rawrr closes this for direct analysis pipelines: "A library providing raw data reading would finally close the gap and facilitate modular end-to-end analysis pipeline development in R"
- [discussion] Windows decimal symbol configuration requirement: "On Windows, the decimal symbol has to be configured as a '.'!"
- [methods] iRT regression model fitting on known peptide standards: "Fit intensity traces stored in rawrrChromatogram objects to extract retention times at maximum intensity (rtFittedAPEX)."
Is this your skill, or is something wrong with this listing? Request removal or report an issue. Author removals are honored within 72 hours.