Skip to content
Back to skills

Internal Standard Ion Selection And Application

ASecurity

Use when after isotope correction has been applied to MSI ion images,

  • 16 stars
  • 0 votes
  • 0 copies
  • 4 views
  • Added September 12, 2026
ai-agentsgogitdatabase

Works with

  • cli

Security analysis

A100/100

Scanned September 12, 2026

npx -y skills add HolobiomicsLab/asb-skill-collections --skill internal-standard-ion-selection-and-application --agent claude-code

Installs into .claude/skills of the current project.

Are you the author of Internal Standard Ion Selection And Application?

Add the live security badge to your README. It updates with every re-scan.

Security grade badge for Internal Standard Ion Selection And Application
[![Security: A — Skills Directory](https://www.skillsdirectory.com/api/skills/holobiomicslab-internal-standard-ion-selection-and-application/badge)](https://www.skillsdirectory.com/skills/holobiomicslab-internal-standard-ion-selection-and-application)

More formats (shields.io, HTML) on the badges page. Keep it an A: scan every change in CI with Pro.

Download with Pro
SKILL.md
---
name: internal-standard-ion-selection-and-application
description: Use when after isotope correction has been applied to MSI ion images,
  when you have sprayed or identified a reference lipid standard of known amount (pmol/mm²)
  and need to normalize target lipid intensities against this standard to remove matrix
  effects and enable cross-pixel and cross-sample.
license: CC-BY-4.0
metadata:
  edam_operation: http://edamontology.org/operation_3215
  edam_topics:
  - http://edamontology.org/topic_0769
  - http://edamontology.org/topic_3520
  tools:
  - LipidQMap
  - Cardinal
  - h5py or h5netcdf
  techniques:
  - MS-imaging
  license_tier: restricted
  provenance_tier: literature
derived_from:
- doi: 10.1101/2025.10.15.682422v1
  title: LipidQMap
evidence_spans:
- LipidQMap writes MSI exports as HDF5 containers
- LipidQMap writes MSI exports as HDF5 containers that follow the [`Cardinal::HDF5`](https://cardinalmsi.org)
  conventions.
claims: []
provenance:
  collection: https://w3id.org/holobiomicslab/asb-skill/collection/metabolomics/v2
  assembled_by: scripts/collect_metabolomics_collection.py
  sources:
  - build: coll_lipidqmap_cq
    doi: 10.1101/2025.10.15.682422v1
    title: LipidQMap
  dedup_kept_from: coll_lipidqmap_cq
schema_version: 0.2.0
attribution:
  generator: AgenticScienceBuilder
  original_doi: 10.1101/2025.10.15.682422v1
  all_source_dois:
  - 10.1101/2025.10.15.682422v1
  zenodo_doi: 10.5281/zenodo.20794027
  curators: []
  promoter: Louis-Félix Nothias
  sponsor: CNRS & Université Côte d'Azur
---

# internal-standard-ion-selection-and-application

> **License: restricted** — no clear open-source license detected for the underlying tool; verify licensing before commercial use or redistribution. <!-- asb-license-banner -->
## Summary

Select and apply user-defined internal standard lipid ions to normalize isotope-corrected ion-image intensities, enabling conversion of raw mass spectrometry imaging signals to quantified values. This skill is essential for converting relative ion intensities into absolute or semi-quantitative lipid abundance estimates across tissue or sample regions.

## When to use

After isotope correction has been applied to MSI ion images, when you have sprayed or identified a reference lipid standard of known amount (pmol/mm²) and need to normalize target lipid intensities against this standard to remove matrix effects and enable cross-pixel and cross-sample comparisons. Use this skill when the goal is to move from raw or isotope-corrected intensity matrices to quantified image data suitable for statistical analysis or biomarker discovery.

## When NOT to use

- Input ion images have not yet undergone isotope correction (use isotope correction skill first)
- No internal standard lipid has been identified, sprayed, or quantified in the sample
- The internal standard intensity is uniformly zero or near-detection limit across most pixels (normalization will produce unreliable or NaN-dominated output)

## Inputs

- HDF5 MSI export file with image_type='isotope' (isotope-corrected ion images)
- User-defined internal standard specification (feature ID or name from database)
- Target lipid feature ID or name to be quantified
- featureData table with feature_id column for standard and target lookup

## Outputs

- HDF5 MSI export file with image_type='quant' containing normalized quantified intensity matrix
- Updated featureData metadata reflecting single quantified target lipid
- Quantified ion image displayable in LipidQMap with Q (quantitative) view toggle

## How to apply

Identify the internal standard lipid species in the sample database by matching either the sample identifier or feature name; retrieve its isotope-corrected intensity row from the HDF5 spectraData/intensity matrix. For each target lipid of interest, retrieve its corresponding intensity row. Perform element-wise division of the target lipid intensity vector by the internal standard intensity vector to produce normalized quantitation. Handle division-by-zero pixels (where standard intensity ≤ 0) by masking them as NaN or zero. Export the normalized matrix to a new HDF5 file with root attribute image_type set to 'quant' and updated featureData metadata following Cardinal::HDF5 conventions. Verify that the quantified image shows spatial patterns consistent with the isotope-corrected view but with reduced pixel-to-pixel noise and improved comparability across regions.

## Related tools

- **LipidQMap** (Primary GUI application for selecting internal standards from database, toggling between raw/isotope-corrected/quantified views, and exporting quantified images) — https://github.com/swinnenteam/LipidQMap
- **Cardinal** (Defines HDF5 container conventions for MSI exports; quantified output must follow Cardinal::HDF5 standard) — https://cardinalmsi.org
- **h5py or h5netcdf** (HDF5 reader/writer library for programmatic access to isotope-corrected intensity matrix and metadata; required for scripted quantification workflows)

## Examples

```
In LipidQMap: (1) Import imzML file with isotope correction enabled. (2) Click on target lipid in Species table. (3) In Settings, confirm the database row 'IS' column points to a standard with 'Is standard'=TRUE and 'Standard amount (pmol/mm2)' populated. (4) Click Settings → select internal standard from dropdown. (5) Press 'Q' to toggle to Quantitative view. (6) Click 'Save images' and select 'Save Quantitative images'.
```

## Evaluation signals

- Output HDF5 file root attribute image_type equals 'quant' (not 'isotope' or 'raw')
- Output quantified intensity matrix shape matches input (n_features × n_pixels) with float32 dtype
- Pixels with zero or negative internal standard intensity are masked as NaN or zero (no inf or invalid values)
- Quantified image view (Q toggle in LipidQMap) displays smooth spatial patterns without extreme outliers caused by low-intensity standard pixels
- featureData metadata in output HDF5 reflects single target lipid (not all original features) with quantitation method documented in file attributes

## Limitations

- Internal standard must be present in the sample at detectable intensity across representative tissue regions; sparse or highly variable standard signal will produce unreliable quantitation
- Quantitation is relative to the chosen internal standard and does not yield absolute abundance without independent standard calibration
- Division-by-zero handling (masking as NaN) reduces effective pixel count in quantified image; high frequency of masked pixels indicates poor internal standard coverage
- Database column 'IS' (internal standard ID reference) must be correctly populated for each target lipid, or standard lookup will fail

## Evidence

- [intro] LipidQMap performs quantitation based on user-defined internal standards: "Performs quantitation based on user defined internal standards."
- [other] Isotope-corrected intensity matrix from HDF5 spectraData/intensity: "Extract the isotope-corrected intensity matrix from spectraData/intensity (shape: n_features × n_pixels, float32)."
- [other] Feature lookup by matching feature_id against user definition: "Locate the internal standard feature in featureData by matching feature_id against the user definition and retrieve its intensity row."
- [other] Element-wise division with NaN/zero masking for zero-intensity pixels: "Divide the target lipid intensity vector element-wise by the internal standard intensity vector (normalized quantitation); handle division-by-zero pixels (standard intensity ≤ 0) by masking them as"
- [other] Output HDF5 follows Cardinal::HDF5 conventions with image_type='quant': "Write the quantified intensity matrix to a new HDF5 file following Cardinal::HDF5 conventions, updating root attribute image_type to 'quant'"
- [readme] Toggle view in LipidQMap between raw, isotope-corrected, and quantified images: "Can toggle view between raw, isotope corrected and quantified images."
- [readme] Database column 'IS' specifies internal standard for quantitation: "**IS**: ID of the standard species that should be used for the quantitation."

Attribution

Is this your skill, or is something wrong with this listing? Request removal or report an issue. Author removals are honored within 72 hours.

View all in ai-agents →

Comments

Loading comments…

Newsletter

Get the weekly list of safe new skills, and the ones we pulled.